Key features
- 200 µm, 300 µm, 400 µm, or 600 µm hard clad silica fibers
- cleaved distal end
- customer’s handpiece
- SMA connector
Fiber optic probes for frontal emission
We offer fiber optic probes with a cleaved end for frontal emission applications. The main applications of such probes are Endovenous Laser Therapy (EVLT), urology, podology, and laser surgery.
FIBERS
Our fibers are USP Class VI certified, non-toxic and biocompatible. Our standard products use either 200 µm, 300 µm, 400 µm or 600 µm core HCS fibers in order to suit any medical procedures and types of lasers.
CONNECTOR
The probe is equipped with the OneShot SMA connector manufactured by SEDI-ATI. This connector is specially designed for disposable probes where cost issue is important. The assembly technique avoids any stress on the fiber itself.
DISTAL END
The simplest design is a perpendicular cleave of the fiber end. To avoid the aggressive sharp end, SEDI-ATI can offer upon request a so-called a-traumatic design where the fiber distal end is fused into a half sphere or a sphere.
MANUFACTURING
Biocompatible materials.
Packaged, stored and used in a clean environment.
Production workshop certified ISO 9001 & ISO 13485.
CE certification according to Directive 93/42, CE number: CE 0476.
Sold sterilized in double bag.
QUALIFICATION / TEST
Internal Qualification Report (RQ091022): 20 testing autoclave cycles.
Test Report n°100210: OneShot Ball Tip, Lateral stress test to failure.
-
MD-OS-200-CL
Fiber typehard clad silica 200 µm core diameterDistal endflat cleavedFiber tip diameter0,63 mmFiber length2.8 mCatheter compatibility2,25FR to 3,12FR depending on the fiber typeHand piececustomer's handpieceConnectorSMA connector -
MD-OS-300-CL
Fiber typehard clad silica 300 µm core diameterDistal endflat cleavedFiber tip diameter0,63 mmFiber length2.8 mCatheter compatibility2,25FR to 3,12FR depending on the fiber typeHand piececustomer's handpieceConnectorSMA connector -
MD-OS-400-CL
Fiber typehard clad silica 400 µm core diameterDistal endflat cleavedFiber tip diameter0,63 mmFiber length2.8 mCatheter compatibility2,25FR to 3,12FR depending on the fiber typeHand piececustomer's handpieceConnectorSMA connector -
MD-OS-600-CL
Fiber typehard clad silica 600 µm core diameterDistal endflat cleavedFiber tip diameter0,63 mmFiber length2.8 mCatheter compatibility2,25FR to 3,12FR depending on the fiber typeHand piececustomer's handpieceConnectorSMA connector